LxT Manual
Acoustic Calibration
7-11
You can simply verify the calibration
by selecting
No
. Also, if no
significant changes are seen you may
choose to answer
No
. However, to
create an historical record for it, or
to export post-calibration data, you
must save it.
When the calibration is completed, select
Yes
to save the
results of the calibration. Select
No
to cancel the results of
the calibration.
Warning Messages
After selecting
Yes
to save the results of the calibration, two
warning messages may appear.
Outside Range of Normal Sensitivity
When the results of the calibration correspond to a
sensitivity greater than 3 dB outside the range of the nominal
sensitivity for that microphone, the message shown in
FIGURE 7-6 appears.
FIGURE 7-6 Outside Normal Sensitivity Range
> 0.5 dB From Previous Calibration Result
When the results of the calibration indicate a change in
sensitivity greater than 0.5 dB from the previous calibration
results, the message shown in FIGURE 7-7 appears.
FIGURE 7-7 0.5 dB From Previous Calibration
Result
Calibration Results
When the calibration results are saved, the
History
tab, as
shown in Figure 7-2, is updated. The parameters for this
most recent calibration appear at the top of the list.
Summary of Contents for SoundTrack LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...