Lake Shore Model 340 Temperature Controller User’s Manual
9-26
Remote Operation
ANALOG
Configure Analog Output Parameters
Input:
ANALOG <output>, [<bipolar enable>], [<mode>], [<input>], [<source>],
[<high value>], [<low value>], [<manual value>]
Returned:
Nothing
Remarks:
Configures the analog output parameters.
<output>
Specifies which analog output to configure.
<bipolar enable>
Specifies whether the analog output is positive output only or bipolar.
<mode>
Specifies data the analog output monitors. Valid entries:
0
= off,
1
= input,
2
= manual,
3
= loop. Loop is valid only if
<output>
=
2
.
<input>
Specifies which input to monitor if
<mode>
=
1
.
<source>
Specifies input data. Valid entries:
1
= kelvin,
2
= Celsius,
3
= sensor units,
4
= linear equation.
<high value>
If
<mode>
is
1
, this parameter represents the data at which the
analog output r100% output.
<low value>
If
<mode>
is
1
, this parameter represents the data at which the analog
output reaches -100% output if bipolar, or 0% output if positive only.
<manual value>
If
<mode>
is
2
, this parameter is the output of the analog output.
Example: ANALOG 1, 1, 2, , , , ,-25.5[term] -
Sets analog output 1 to manual mode with -25.5% output
(-2.55 V).
ANALOG 2, 0, 1, A, 1, 100.0, 0.0[term] -
Sets analog output 2 to monitor Input A kelvin
reading with 100.0 K at +100% output (+10.0 V) and 0.0 K at 0% output (0.0 V).
ANALOG?
Query Analog Output Parameters
Input: ANALOG?
<output>
Returned:
<bipolar enable>, <mode>, <input>, <source>, <high value>, <low value>, <manual value>
Format:
n,n,ann,n,
±
nnn.nnnE
±
n,
±
nnn.nnnE
±
n,
±
nnn.n[term]
Remarks:
Returns the parameters of an analog output. See the
ANALOG
command for parameter
descriptions.
<output>
specifies which analog output to query.
AOUT?
Query Analog Output Data
Input: AOUT?
<output>
Returned: <analog output >
.
Format:
±
nnn.n[term]
Remarks:
Returns the percentage of output of an analog output.
<output>
specifies which analog
output data to query.
BEEP
Configure System Beeper Parameters
Input: BEEP
<off/on>
Returned:
Nothing
Remarks:
Enables or disables system beeper sound when an alarm condition is met.
<off/on>
disables/enables the beeper.
BEEP?
Query System Beeper Parameters
Input: BEEP?
Returned: <off/on>
. Format:
n[term]
Remarks:
Returns system beeper parameters. See the
BEEP
command for parameter descriptions.
BEEPST?
Query System Beeper Status
Input: BEEPST?
Returned: <beeper status>
. Format:
n[term]
Remarks:
Returns the current condition of the system beeper.