INSTRUCTIONS FOR PROPER USE
FOREHEAD TEMPERATURE RECORDING (Body)
Attention!
In order to ensure accurate recording of the temperature, clean the probe in its entirety with a cotton swab
moistened with alcohol after each use. Make sure that the probe is completely dry before use.
IMPROPER CLEANING OF THE PROBE CAN CAUSE UNRELIABLE TEMPERATURE RECORDINGS
1) Clean and dry the forehead where you will be positioning the probe.
2) Point the probe at the centre of the forehead just above the space between the eyebrows, at a distance of
approximately 1-5 cm.
3) Press the "O/I" key, all the function symbols (self-test) will appear on the display; immediately after the
appliance emits an acoustic signal and the display immediately shows the measured value.
The green display indicates a normal temperature (between 34 and 37.3°C/93.2-99.1°F), the yellow colour
indicates the presence of fever (temperature between 37.4 and 38°C/99.3-100.4°F), the red colour indicates
high fever (with temperature between 38.1 and 43°C/100.6-109.4°F).
The display shows the symbol
when the measured body temperature is below 37.4°C (99.3°F) and the
symbol
when it is higher.
When the temperature exceeds 37.4°C the thermometer emits 4 short beeps.
The temperature shown on the display is equivalent to the underarm temperature.
4) The appliance switches off automatically after approximately 18 seconds.
Wait at least one second between one measurement and the next.
After 5 consecutive measurements, wait at least 30 seconds before performing a new recording.
RECORDING OF THE TEMPERATURE OF THE ENVIRONMENT, SURFACES AND WARM LIQUIDS (Surface)
This thermometer allows recording of the temperature of the environment, of surfaces or of hot liquids (such
as milk, water, ...).
1) Turn on the thermometer by pressing the "MODE" key. The display shows "37.0°C (Body)".
2) Press the "MODE" key again: "Surface" appears on the display.
3) To measure the temperature, bring the thermometer close to the relevant area (water, milk container, ...) at a
distance of approximately 1-5 cm, press the "O/I" key, the appliance emits an acoustic signal and the display
immediately shows the measured value.
4) The appliance switches off automatically after approximately 18 seconds of non-use. When it is switched on
again, the display shows the forehead temperature measurement function (Body).
Wait at least one second between one measurement and the next.
After 5 consecutive measurements, wait at least 30 seconds before performing a new recording.
N.B.:
Never immerse the probe in any kind of liquid.
SILENT MEASUREMENT MODE
This thermometer records the temperature in silent mode making it ideal for measurements on children
even during sleep.
1) Turn on the thermometer by pressing the "SET" key: the display shows the forehead temperature
measurement function (Body).
2) Press the "SET" key again: the display shows the word "OFF" and the flashing symbol
in
succession.
3) Proceed with the measurement as described in the previous paragraphs.
4) To activate the acoustic signal, turn on the thermometer by pressing the "SET" key: the display shows the
forehead temperature measurement function (Body).
5) Press the "SET" key again: the thermometer emits an acoustic signal and the display shows successively
the word "On" and the flashing symbol
that remains displayed during the next measurement.
MEMORY FUNCTION
This appliance stores the last 50 measurements to assess any changes. Readings are automatically stored.
Once 50 readings have been exceeded, the oldest data is automatically deleted.
To recall the stored readings, press the "MEMO" key. Each press of the key scrolls the data in the memory:
number 1 refers to the most recent reading, 50 to the oldest one. The appliance switches off automatically
after approximately 18 seconds of non-use.
It is possible to delete the stored data: when the appliance is on, keep the "MEMO" key pressed until
the display shows "CLr", the thermometer emits an acoustic signal and the display shows the forehead
temperature measurement function (Body).
THERMOMETER SETTINGS
• MEASURE UNIT - "F1" FUNCTION
This thermometer can measure temperature in Celsius and Fahrenheit.
1) Turn on the thermometer by pressing the "SET" key: the display shows 37.0°C. Keep the key pressed until the
display shows "F1". When the key is released, the display shows "Unit".
2) Press the "MEMO" or "MODE" key to select the unit of measurement and press the "SET" key to confirm it: the
display shows "F2".
• "HIGH FEVER ALARM" – "F2” FUNCTION
It is possible to set the minimum temperature signalled as "high fever alarm". The thermometer is set to signal this