INSTALLATION

Open the illumination house to install the mercury lamp. Loosen the 

thumbscrew indicated by the arrow and gently  pull out lamp door from lamp 

housing. 

To avoid any damage to the mercury vapor lamp, a plastic rod (1) is fitted in 

place  of lamp during transit. Remove this rod by loosening the two screws 

indicated by the two arrows A and B shown in Fig.

 

 6.

Mount the mercury vapor lamp in place of plastic rod and tighten the screws 

to secure the lamp. 

in

s

t

a

ll

 as s

hown in Fig

. 6.A.

 

Avoid touching the lamp in order to prevent leaving finger prints on the 

glass. If 

there is any residue on the lamp's surface, gently wipe it off with 

soft tissue paper.

Install the power cable coming from the illumination house to the power unit 

in

d

i

ca

te

d by a

rrow

 A

 in Fig

. 7.

(1).  Connect the power cord to the AC Inlet

 in

d

i

ca

te

d by a

rrow

 

in Fig

. 7.

Fluorescence Attachment User Guide 

5

Fig. 7

Fig. 6

B

 

A

Fig

. 6A

 

Summary of Contents for TCM 400 FLR

Page 1: ...ting instrument understanding this manual completely before use is highly recommended User Manual TCM 400FLR Inverted Microscopy R 7127000 795 _________________________________________________________...

Page 2: ...TCM400Flr...

Page 3: ...UBLESHOOTING 8 10 TECHNICAL SPECIFICATION 9 8 OBSERVING A SPECIMEN 7 7 CENTERING THE LIGHT BEAM 7 6 MAINTAINING OPERATING LIFE 6 5 INSTALLATION 4 4 SAFETY INSTRUCTIONS 3 3 UNPACKING 2 2 PRINCIPLE 1 1...

Page 4: ...PACKAGING CONTENTS 1 FL attachment in 2 parts 2 2 Filter cube 3 Power Supply 4 Power cord 5 UV protection shield Fluorescence Attachment User Guide 2...

Page 5: ...h to OFF O and unplug the lamp housing s connecting cord plug from the output connector on the power supply unit Wait at least 10 minutes or until the lamp housing cools down before any changes are ma...

Page 6: ...ssure by thumbs on the inner sides of the panel 3 4 to push it out Refer to gure 1 and 2 2 Join the Optical assembly and illumination assembly to complete the FL attachment Guide the illumination asse...

Page 7: ...d B shown in Fig 6 Mount the mercury vapor lamp in place of plastic rod and tighten the screws to secure the lamp install as shown in Fig 6 A Avoid touching the lamp in order to prevent leaving nger p...

Page 8: ...ited at least 5 minutes required to stabilize light To avoid shortening the power supply unit s operating life avoid turning power ON while the lamp is not mounted Avoid turning power ON and OFF withi...

Page 9: ...Adjust the beam position to the center of the eld of view asshown in Fig 9 A OBSERVING A SPECIMEN Fluorescence Attachment User Guide 7 1 A B Fig 9 Turn the power supply ON Allow the lamp at least 5 mi...

Page 10: ...cus deviates from correct Adjust the condenser knob position This process is shown on Page 7 4 Mercury lamp is not illuminating Some mercury lamps may not ignite Press the Trigger button provided afte...

Page 11: ...85 nm onwards Dichroic mirror Transmission 93 at wavelength 540 nm onwards Transmission 0 at wavelength 400 to 530 nm Barrier lter Transmission 95 at wavelength 535 nm onwards Transmission 0 at wavele...

Page 12: ...400 FLR are registered trademarks of Labo America Inc With a policy of continuous development Labo America Inc reserves the right to change design and speci cations without prior notice 2011 Labo Amer...

Reviews: