CONTENTS
9
TROUBLESHOOTING
8
10
TECHNICAL SPECIFICATION
OBSERVING A SPECIMEN
7
CENTERING THE LIGHT BEAM
6
MAINTAINING OPERATING LIFE
5
INSTALLATION
4
SAFETY INSTRUCTIONS
3
UNPACKING
2
PRINCIPLE
1
INTRODUCTION
Page 1: ...ting instrument understanding this manual completely before use is highly recommended User Manual TCM 400FLR Inverted Microscopy R 7127000 795 _________________________________________________________...
Page 2: ...TCM400Flr...
Page 3: ...UBLESHOOTING 8 10 TECHNICAL SPECIFICATION 9 8 OBSERVING A SPECIMEN 7 7 CENTERING THE LIGHT BEAM 7 6 MAINTAINING OPERATING LIFE 6 5 INSTALLATION 4 4 SAFETY INSTRUCTIONS 3 3 UNPACKING 2 2 PRINCIPLE 1 1...
Page 4: ...PACKAGING CONTENTS 1 FL attachment in 2 parts 2 2 Filter cube 3 Power Supply 4 Power cord 5 UV protection shield Fluorescence Attachment User Guide 2...
Page 5: ...h to OFF O and unplug the lamp housing s connecting cord plug from the output connector on the power supply unit Wait at least 10 minutes or until the lamp housing cools down before any changes are ma...
Page 6: ...ssure by thumbs on the inner sides of the panel 3 4 to push it out Refer to gure 1 and 2 2 Join the Optical assembly and illumination assembly to complete the FL attachment Guide the illumination asse...
Page 7: ...d B shown in Fig 6 Mount the mercury vapor lamp in place of plastic rod and tighten the screws to secure the lamp install as shown in Fig 6 A Avoid touching the lamp in order to prevent leaving nger p...
Page 8: ...ited at least 5 minutes required to stabilize light To avoid shortening the power supply unit s operating life avoid turning power ON while the lamp is not mounted Avoid turning power ON and OFF withi...
Page 9: ...Adjust the beam position to the center of the eld of view asshown in Fig 9 A OBSERVING A SPECIMEN Fluorescence Attachment User Guide 7 1 A B Fig 9 Turn the power supply ON Allow the lamp at least 5 mi...
Page 10: ...cus deviates from correct Adjust the condenser knob position This process is shown on Page 7 4 Mercury lamp is not illuminating Some mercury lamps may not ignite Press the Trigger button provided afte...
Page 11: ...85 nm onwards Dichroic mirror Transmission 93 at wavelength 540 nm onwards Transmission 0 at wavelength 400 to 530 nm Barrier lter Transmission 95 at wavelength 535 nm onwards Transmission 0 at wavele...
Page 12: ...400 FLR are registered trademarks of Labo America Inc With a policy of continuous development Labo America Inc reserves the right to change design and speci cations without prior notice 2011 Labo Amer...