Labomed TCM 400 FLR User Manual Download Page 10

TROUBLESHOOTING 

Fluorescence Attachment User Guide 

8

Problem 

Cause 

Remedy

Optical System

1) 

The mercury lamp lights up, 

Filter slider is not engaged properly 

Adjust filter to required setting.

but field of view remains dark 

in light path. 

2)

Image is unclear, blurred or

a)

Objectives and/or filters are dirty.

Clean them with tissue/lens paper.

has poor contrast.

b)

Field iris diaphragm is not

Adjust field iris diaphragm so 

fully opened.

that the image circumscribes  

the field of view.

3)

Field of view is vignetted or

a)

Objective is not engaged properly

Be sure when rotating nosepiece

it is not evenly illuminated.

in light path.

that objective clicks into position.

b)

Filter slider is not engaged properly

Engage required filter 

in light path.

properly in light path.

c)

Mercury lamp is not centered

Center mercury lamp correctly. 

properly.

This process is shown on Page 7

d)

Beam focus deviates from correct

Adjust the condenser knob. 

position.

This process is shown on Page 7

4)

Mercury lamp is not illuminating

Some mercury lamps may not ignite 

Press the Trigger button provided 

after switching power supply ON.

the first time the power supply is  

on the front panel of power supply

turned ON. 

 unit and wait for lamp to ignite. 

Electrical System

1)

Lamp housing / Main switch

a)

Power cord is not connected

Make sure power cord is firmly

indicator does not light up. 

properly.

in place.

b)

The fuse is blown.

Replace with fuse of same type  

and rating.

2)

Main switch indicator lights up,

a)

Connector cable is not engaged

Verify connection is secure.

yet mercury lamp remains OFF.

properly.

b)

Mercury lamp is not mounted.

Install mercury lamp as 

indicated on Page 5

3)

Light demonstrates partial

a)

Sufficient time has not been

Allow  at least 5 minutes after

flickering.

allowed for the lamp to fully

turning power supply ON for the

illuminate.

lamp to fully illuminate. If flickering 

persist press Trigger button.

b)

Lamp's service life has expired.

Replace the mercury lamp 

Summary of Contents for TCM 400 FLR

Page 1: ...ting instrument understanding this manual completely before use is highly recommended User Manual TCM 400FLR Inverted Microscopy R 7127000 795 _________________________________________________________...

Page 2: ...TCM400Flr...

Page 3: ...UBLESHOOTING 8 10 TECHNICAL SPECIFICATION 9 8 OBSERVING A SPECIMEN 7 7 CENTERING THE LIGHT BEAM 7 6 MAINTAINING OPERATING LIFE 6 5 INSTALLATION 4 4 SAFETY INSTRUCTIONS 3 3 UNPACKING 2 2 PRINCIPLE 1 1...

Page 4: ...PACKAGING CONTENTS 1 FL attachment in 2 parts 2 2 Filter cube 3 Power Supply 4 Power cord 5 UV protection shield Fluorescence Attachment User Guide 2...

Page 5: ...h to OFF O and unplug the lamp housing s connecting cord plug from the output connector on the power supply unit Wait at least 10 minutes or until the lamp housing cools down before any changes are ma...

Page 6: ...ssure by thumbs on the inner sides of the panel 3 4 to push it out Refer to gure 1 and 2 2 Join the Optical assembly and illumination assembly to complete the FL attachment Guide the illumination asse...

Page 7: ...d B shown in Fig 6 Mount the mercury vapor lamp in place of plastic rod and tighten the screws to secure the lamp install as shown in Fig 6 A Avoid touching the lamp in order to prevent leaving nger p...

Page 8: ...ited at least 5 minutes required to stabilize light To avoid shortening the power supply unit s operating life avoid turning power ON while the lamp is not mounted Avoid turning power ON and OFF withi...

Page 9: ...Adjust the beam position to the center of the eld of view asshown in Fig 9 A OBSERVING A SPECIMEN Fluorescence Attachment User Guide 7 1 A B Fig 9 Turn the power supply ON Allow the lamp at least 5 mi...

Page 10: ...cus deviates from correct Adjust the condenser knob position This process is shown on Page 7 4 Mercury lamp is not illuminating Some mercury lamps may not ignite Press the Trigger button provided afte...

Page 11: ...85 nm onwards Dichroic mirror Transmission 93 at wavelength 540 nm onwards Transmission 0 at wavelength 400 to 530 nm Barrier lter Transmission 95 at wavelength 535 nm onwards Transmission 0 at wavele...

Page 12: ...400 FLR are registered trademarks of Labo America Inc With a policy of continuous development Labo America Inc reserves the right to change design and speci cations without prior notice 2011 Labo Amer...

Reviews: