SAFETY INSTRUCTIONS 

Fluorescent attachment is manufactured according to CE safety norms and regulations.

Fluorescent attachment is intended for use only as prescribed in this manual.

Make sure the main switch on the power supply unit is set to OFF (O) before supplying power to the unit.

The power supply unit contains high voltage components. Do not attempt to disassemble the unit when turned ON 

(I).

To avoid injury, do not remove the lamp housing while the lamp is lit. Also, do not turn the lamp ON while the lamp 

housing is removed to avoid damaging the instrument.

Before opening lamp housing for replacement of the lamp, set the main switch to OFF (O) and unplug the lamp 

housing's connecting cord plug from the output connector on the power supply unit. Wait at least 10 minutes or until 

the lamp housing cools down before any changes are made.

Do not open the lamp housing while the system is turned ON, for any user contact with internal parts may cause 

injury.

A used mercury burner should be disposed of in compliance with the ordinances or regulations of your national or 

local government.

The manufacturer will not accept any liability for damage caused by any user following unsafe practice of the 

instrument.

Fluorescence Attachment User Guide 

3

Avoid using the lamp beyond its service life. The mercury lamp seals high pressure gasses. If the lamp is used beyond its service life, 
stress may accumulate inside the lamp and in the worst but very rare case, the lamp may burst. 

Summary of Contents for TCM 400 FLR

Page 1: ...ting instrument understanding this manual completely before use is highly recommended User Manual TCM 400FLR Inverted Microscopy R 7127000 795 _________________________________________________________...

Page 2: ...TCM400Flr...

Page 3: ...UBLESHOOTING 8 10 TECHNICAL SPECIFICATION 9 8 OBSERVING A SPECIMEN 7 7 CENTERING THE LIGHT BEAM 7 6 MAINTAINING OPERATING LIFE 6 5 INSTALLATION 4 4 SAFETY INSTRUCTIONS 3 3 UNPACKING 2 2 PRINCIPLE 1 1...

Page 4: ...PACKAGING CONTENTS 1 FL attachment in 2 parts 2 2 Filter cube 3 Power Supply 4 Power cord 5 UV protection shield Fluorescence Attachment User Guide 2...

Page 5: ...h to OFF O and unplug the lamp housing s connecting cord plug from the output connector on the power supply unit Wait at least 10 minutes or until the lamp housing cools down before any changes are ma...

Page 6: ...ssure by thumbs on the inner sides of the panel 3 4 to push it out Refer to gure 1 and 2 2 Join the Optical assembly and illumination assembly to complete the FL attachment Guide the illumination asse...

Page 7: ...d B shown in Fig 6 Mount the mercury vapor lamp in place of plastic rod and tighten the screws to secure the lamp install as shown in Fig 6 A Avoid touching the lamp in order to prevent leaving nger p...

Page 8: ...ited at least 5 minutes required to stabilize light To avoid shortening the power supply unit s operating life avoid turning power ON while the lamp is not mounted Avoid turning power ON and OFF withi...

Page 9: ...Adjust the beam position to the center of the eld of view asshown in Fig 9 A OBSERVING A SPECIMEN Fluorescence Attachment User Guide 7 1 A B Fig 9 Turn the power supply ON Allow the lamp at least 5 mi...

Page 10: ...cus deviates from correct Adjust the condenser knob position This process is shown on Page 7 4 Mercury lamp is not illuminating Some mercury lamps may not ignite Press the Trigger button provided afte...

Page 11: ...85 nm onwards Dichroic mirror Transmission 93 at wavelength 540 nm onwards Transmission 0 at wavelength 400 to 530 nm Barrier lter Transmission 95 at wavelength 535 nm onwards Transmission 0 at wavele...