INSTALLING THE FILTER
B & G
Select the filter to be used on the fluorescence attachment.
Engage the slider to the filter house and gently slide it in.
It will
stop at the
I
i
n
d
e
x
position.
Ensure that the power supply unit has been connected to proper power supply voltage.
Set the main switch on the power supply unit to (
|
). After the lamp is ignited, at least 5 minutes required
to
s
t
ab
ili
z
e
light
.
.
To avoid shortening the power supply unit's operating life, avoid turning power ON while the lamp is not mounted
Avoid turning power ON and OFF within short intervals for this may reduce working life of the power supply.
After the lamp is turned OFF, it should not be re-ignited before the mercury vapors cool and condense. Wait
approximately 10 minutes before turning power to the fluorescent attachment O
N again.
MAINTAINING OPERATING LIFE
Fluorescence Attachment User Guide
6
Fig. 8