© 2018 Corning Incorporated
35
Thermal Chamber delay
With
keys choose CHAMBER DEL.
Press SET -
appears.
With
keys select time value.
Set demanded value by pressing
.
Press SET.
Press BACK.
Thermal chamber (Constant temperature in chamber without centrifuging)
Via
keys choose THERM.
CHAMB.
Press SET (to switch off/on).
Via
keys mark value of temperature.
Press SET -
appears.
Via
keys SET demanded temperature.
Press BACK.
Activation of thermal chamber is delayed
in accordance with information content in
chapter 8.5 Thermal chamber delay.
When THERMAL CHAMBER function is
activated,
symbol is blinking on the
screen.
Changing temperature from the main
screen is not possible.
Opening cover terminates THERM.
CHAMB. function (closing cover back turns
it on).
If THERMAL CHAMBER is turned on (in PARAM fold) and centrifugation completes,
THERMAL CHAMBER will activate itself.
THEMRAL CHAMBER can be only activated when any other program is not running.