© 2018 Corning Incorporated
34
When density is changed, symbol
is visible on the screen.
Increasing density of the sample above
1,2 g/cm
3
(and limiting of the maximum speed
of centrifuging resulting from it) applies until
switching off power supply of the centrifuge
or setting the device back to 1,2 g/cm
3
.
Increasing the density reduces the maximum
speed of the rotor.
Temperature offset
With
keys select TEMP. OFFSET.
Press SET-
appears.
Use the
keys to select the
difference between the temperature
that the cooling system will aim for, and
set temperature.
Confirm selection by pressing SET.
Press BACK
Attention!
The use of the offset can not extend the
temperature range achieved by the
centrifuge.
Function description
At a set temperature of 20
o
C and the set offset
value equal to -5
o
C, cooling system will
actually strive to reach 15
o
C. With a setpoint
temperature of 20
o
C and a set offset value of
5
o
C the system will actually try to reach 25
o
C.
The temperature displayed on the main
screen is corrected for offset value.
Offset can be selected range from -20
o
C to
20
o
C.
Activation of the function is signaled on the
main screen with
or
depending on
the offset value sign.