Screen messages that may occur during operation.
MESSAGE
EXPLANATION
"SPEED OF ROTOR"
"IDENTIFICATION <> 90 RPM"
SPEED OF ROTOR IDENTIFICATION <> 90 RPM
"IMBALANCE FAST STOP!"
"PLEASE REMOVE CAUSE"
"THEN RESTART"
UNBALANCE DETECTED
"NO ROTOR OR IDENTIFICATION"
"SENSOR DAMAGED!"
ERROR OF ROTOR IDENTIFICATION {LIMIT OF 6SEC. IS OVER}
"INCORRECT ROTOR NUMBER!"
ROTOR’S IS NOT CORRECT
"WRONG DIRECTION OF ROTATION"
"OR UNKNOWN ROTOR!"
WRONG DIRECTION OF ROTATION / UNKNOWN ROTOR
"PLEASE CLOSE THE LID"
"HAND!"
CLOSING THE LID MANUALLY
"ROTOR STOPPING!"
"Please wait..."
INITIALIZING AFTER MAINS FAILURE WITH ROTATING
ROTOR
" CYCLE'S ABORTED!"
CENTRIFUGATION ENDED BECAUSE STOP WAS PRESSED
" CYCLE'S FINISHED"
CENTRIFUGING ENDED {WITHOUT ERRORS}
Emergency messages
In case of emergency messages (centrifuge is not working properly) contact the manufacturer's authorized service centre.
MESSAGE
"OVERHEATING MOTOR!"
“INVERTER ERROR!"
"INVERTER SERIAL BUS ERROR!"
"TEMPERATURE SENSOR ERROR"
"OPENING COVER in RUN!"
"SPEED METER ERROR"
"I2C BUS ERROR"
"OVERHEATING CENTRIFUGE!"
"ROTOR OVERSPEED!"
"COVER LOCK MALFUNCTION!"
"WORKING 2000 HOURS:”
"CALL SERVICE FOR"