© 2018 Corning Incorporated
37
START DELAY / OF TIME function cannot be run when START DELAY / OF TEMP. is
activated.
Start delay – of temperature
Start centrifuging since preselected
delay is reached.
START DELAY / OF TEMP
Via
keys select START DELAY.
Press SET.
Via
keys select OF TEMP.
Press SET.
Via
keys select temperature zone.
Press SET-
appears.
Via
keys set demanded value.
Press SET.
Press BACK.
When START DELAY – OF TEMP is turned on,
symbol is visible on the screen.
When the function is active, the speed can be reduced to the optimum values for the FAST
COOL function, when the set speed is lower than the optimum value, the rotor rotates at the
set speed.
It is possible to exit START DELAY –of temperature mode at any time by pressing STOP key.
START DELAY / OF TEMP. function cannot be run when START DELAY / OF TIME is activated.