Appendix C: RapidVap Specifications
Product Service 1-800-522-7658
61
Original instructions
Solvent:
Methylene
Chloride
Ethyl
Acetate
Hexanes
Acetone
Toluene
Water
Block Preheat &
Block Set Point
Temperature:
40°C
75°C
70°C
55°C
100°C
95°C
Starting Volume
450 ml
58
56
33
68
75
420
400
53
48
29
61
63
352
350
48
40
24
54
52
321
300
41
35
21
47
45
266
250
35
31
19
43
40
230
200
30
26
16
38
35
193
150
25
22
14
34
28
153
100
19
17
11
29
22
120
50
10
9
7
16
15
68
Table B
Evaporation Time in Upper Portion of Sample Vial
(600 ml Sample Tubes) (Time in Minutes)
N
Number of samples: 8
Nitrogen pressure: 10 psi at volumes 0 thru 300 ml
(see Note 1)
5 psi at volumes above 300 thru 350 ml
4 psi at volumes above 350 thru 400 ml
2 psi at volumes above 400 thru 450 ml
Vortex speed: 70% at volumes 0 thru 300 ml
(see Note 1) 36% at volumes above 300 thru 450 ml
Sample tubes: Not covered with caps