Appendix C: RapidVap Specifications
Product Service 1-800-522-7658
58
Original instructions
Evaporation Rates
RapidVap Vacuum Models
The following tables show the approximate time to evaporate various solvents in
75mm and 16mm sample tubes when the sample block is fully loaded. Other
sample blocks are available for different size tubes. If they are used or if a
different number of samples is run at one time, the evaporation times can be
estimated from these charts or determined experimentally.
Evaporation Time
– 75 mm dia. Sample Tube
(Time in Minutes)
Solvent:
Methylene
Chloride
Toluene
Acetonitrile
Water
Block Preheat &
Block Set Point
Temperature:
50°C
80°C
75°C
80°C
Vacuum:
250 mbar
133 mbar
167 mbar
133 mbar
Starting Volume
300ml
62
90
102
402
250
56
72
84
345
200
49
66
71
289
150
42
57
60
239
100
36
33
49
189
50
22
40
37
95
Number of samples: 8
Vortex speed: 28%
Vacuum pump: Labconco model – 79249-00
Rated free air flow – 88 L/min
Max. attainable vacuum – 31 mbar
V