Appendix C: RapidVap Specifications
Product Service 1-800-522-7658
63
Original instructions
RapidVap N
2
/48 Models
The following tables show the approximate time to evaporate various solvents in
12mm and 20mm sample tubes when the sample block is fully loaded. Other
sample blocks are available for different size tubes. If they are used or if a
different number of samples is run at one time, the evaporation times can be
estimated from these charts or determined experimentally. Evaporation rates may
vary from tube to tube.
Solvent:
Methylene
Chloride
Toluene
Acetonitrile
Methanol
Hexane/
Ethyl Acetate
75:25
Hexane/
Ethyl Acetate
50:50
Block Preheat
& Block Set
Point Temp:
40°C
80°C
75°C
40°C
40°C
40°C
Starting Volume
2 ml
6
8
8
18
5
5
Evaporation Time RapidVap N
2
/48 Models
(12 mm Sample Tubes) (Time in Minutes)
Number of samples: 48
Vortex speed: 100%
Nitrogen pressure: 8 psi
N