Before you install your RapidVap, you need to prepare your site for installation.
Carefully examine the location where you intend to install your RapidVap. You
must be certain that the area is level and of solid construction. In addition, an
exhaust means must be provided and a source of nitrogen must be available for the
RapidVap N
2
. An electrical source must be located near the installation site.
Carefully read this chapter to learn:
the electrical supply requirements.
the exhaust requirements.
the nitrogen supply requirements for the RapidVap N
2
.
the vacuum pump requirements for the RapidVap Vacuum.
Refer to
Appendix C: RapidVap Specifications
for complete RapidVap electrical
and environmental conditions, specifications and requirements.
If the unit is not operated as specified in this manual it may
impair the protection provided by the unit.
Si l'appareil n'est pas utilisé comme spécifié dans ce manuel, il
peut compromettre la protection offerte par l'appareil.
Electrical Requirements
If your RapidVap is a Vacuum model, a dedicated electrical outlet is required.
This outlet requires a 20 Amp circuit breaker or fuse for models rated at 115V (60
Hz). A 10 Amp circuit breaker or fuse is required for models rated at 230V (50/60
Hz). 115V models are equipped with a 20 Amp NEMA 5-20P plug. 230V models
are equipped with a 16 Amp CEE 7/7 plug. It may be necessary to remove the
plug and install a different plug to match the available receptacle.