Chapter 7: Troubleshooting
Product Service 1-800-522-7658
42
Original instructions
DISPLAY
ERROR CODE
CAUSE
CORRECTIVE ACTION
Mem Fail – Push
Prog
Memory failure
Push program button.
New memory IC
chip
Push program button.
Bad memory IC
chip
Call Labconco – Replace IC chip or control
PCB.
Motor Error
Stalled condition
Turn power off. Try to restart after one
minute. Call Labconco if error message
remains.
Excessive speed
detected, i.e. >120%
Turn power off. Try to restart after one
minute. Call Labconco if error message
remains.
Defective motor
Turn power off. Try to restart after one
minute. Call Labconco if error message
remains.
Other corrective actions for potential problems are as follows:
PROBLEM
CAUSE
CORRECTIVE ACTION
Unit will not
operate
Unit not connected to
electrical power
Connect unit to proper electrical
receptacle.
Circuit breaker blown
Correct electrical problem and reset
circuit breaker by pressing button.
Lid open
Close lid.
Excessive
vibration
Sample tubes not
located symmetrically
in block
Reposition sample tubes.
Sample block loose
Tighten all three mounting screws.
Sample block not
installed
Install sample block.
Sample odor in
lab
Vent hose exhausting
into lab area
Redirect hose to fume hood.
Inoperable blower
(RapidVap N
2
)
Replace blower.