Chapter 4: Using Your RapidVap
Product Service 1-800-522-7658
35
Original instructions
#t[xxxx]
Selects the TIME SET POINT of program 9 from 1 to 999
minutes, and 1000 for no stopping (will run continuously), will
get a xxxx;yyyy’\n’ response for confirmation. (xxxx:time
setpoint, yyyy:time left)
#V[xxxx];
Selects the VACUUM SET POINT (if the RapidVap is a
Vacuum model) of program 9 from 1 to 1000 mBar, will get a
xxxx;yyyy’\n’ response for confirmation. (xxxx:vacuum
setpoint, yyyy:actual vacuum)
#H[xxx];
Selects the PRE-HEAT SET POINT of program 9 from 30 to
100 C, will get a xxx;yyy’\n’ response for
confirmation.(xxx:pre-heat setpoint, yyy:actual temperature)
#L[x];
KEY LOCKED when x=1 and UNLOCKED when x=0, will
get a x’\n’ response for confirmation
#N[x]
elects the N2 VALVES (if the RapidVap is an N
2
model) of
program 9 from 0,2,4,6,8, will get a x’\n’ response for
confirmation
#M[x];
x=1 will mute the audio alarm; x=0 will allow the audio alarm
to work normally
#E;
The RapidVap will respond to this command with a numeric
value as follows :’”1"’=indicates an endpoint has been reached;
“0”=endpoint has not been reached
#e;
The RapidVap will respond to this command with a numeric
value as follows: 0=all functions are normal; “1”=lid is open;
“2”=communication problem; “3” and later numbers will be
established later as specific error messages
Note: When controlling the RapidVap using the RS-232 link, all control parameters
will be entered into program 9.
Note: All of the above commands can be sent with no numeric value, that is
‘#[command];’ (without value) to read out the setpoint and actual value without
changing the setpoint.
Note: There is no command for the vacuum release button on the front panel since
this button is momentary in nature – the “#V0000;” command will bleed the
vacuum chamber to atmospheric pressure.
Note: The above commands can be sent at a maximum rate of one command per
second.