Before you install your CentriVap, you need to prepare
your site for installation. Carefully examine the
location where you intend to install your CentriVap.
You must be certain that the area is level and of solid
construction. In addition, an exhaust means must be
provided. An electrical source must be located near the
installation site.
Carefully read this chapter to learn:
•
the electrical supply requirements.
•
the exhaust requirements.
•
the vacuum pump requirements.
Refer to
Appendix C: CentriVap Specifications
for
complete CentriVap electrical and environmental
conditions, specifications and requirements.
The CentriVap has not been evaluated
by an approval agency for the use of
biological, radiotoxins or flammable
liquids or materials.
!