Chapter 4: Using Your CentriVap
Depending on which solvent is used, the volume of the
sample and the system operating parameters, the Cold
Trap may warm up sufficiently to cause the “READY”
indicator to turn off. As the sample reaches
completion, the load on the Cold Trap will decrease and
its temperature will decrease. The “READY” indicator
will once again illuminate.
The LCD display conveys various pieces of
information. When the CentriVap is turned on, the
CentriVap returns to the same mode (“RUN” or
“STOP”) that it was in when the power was turned off.
If the CentriVap was in the “RUN” mode when the
power was turned off, when the power is turned on, the
CentriVap attempts to return to the programmed set
points and continue the run.
Interrupting a Cycle After it
Has Begun
At any time during a run, the cycle may be stopped by
pressing the “STOP” button. This shuts off all
operating functions. After the rotor stops, the lid may
be opened. If it is necessary to re-start the CentriVap,
close the lid and press “RUN.” The CentriVap resumes
operation at the same set point parameters and the timer
continues to count down from the time at which the
CentriVap was stopped.