Appendix C: CentriVap Specifications
Water
bp 100
Tube Size
(ml)
Number
of
Samples
Sample
Size (ml)
Heater
Temp
(C)
Heat
Boost
Vacuum
Pump
Time to
Dry
(min)
Cold
Trap
Rate
Overall
(ml/min)
50 12 25 45
off
Diaphragm
1397
no
0.21
50
12 25 100 off Diaphragm 445 no 0.67
50
12 25 100 on Diaphragm 430 no 0.70
15 18 10 45
off
Diaphragm
841
no
0.19
1.5 132 1 45
off
Diaphragm
427
no
0.30
1.5 132 1 60
off
Diaphragm
303
no
0.44
1.5 132 1 75
off
Diaphragm
196
no
0.67
1.5 132 1 100 off Diaphragm 117 no 1.13
1.5 132 1 100 on Diaphragm 108 no 1.22
50 12 25 45
off
rotary
vane
1002
yes
0.30
50 12 25 100
off
rotary
vane
424
yes
0.70
50 12 25 100
on
rotary
vane
373
yes
0.80
15 18 10 45
off
rotary
vane
565
yes
0.32
15 18 10 60
off
rotary
vane
456
yes
0.39
1.5 132 1 45
off
rotary
vane
299
yes
0.44
1.5 132 1 75
off
rotary
vane
207
yes
0.64
1.5 132 1 100
off
rotary
vane
131
yes
1.01
1.5 132 1 100
on
rotary
vane
122
yes
1.08
Vacuum Pumps:
Diaphragm – Rated at 12 mbar ultimate vacuum and 34 L/min displacement
Rotary vane – Rated at 2.0 x 10
-3
mbar ultimate vacuum and 195 L/min displacement