Chapter 06 – Measuring range 37
10911 V01.10
Note:
The search function in the operating mode of JESS has been deactivated.
6.5.7 LNB current measurement
The measuring receiver measures the power of the direct current that comes from the RF input (e.g. for a LNB supply)
and indicates it in mA in the LNB window. The measuring range is between 0 and 500 mA, the resolution is 1 mA.
6.5.8 DiSEqC Skript
Newer multiswitches for single-cable systems are programmable by DiSEqC-signals (like JULTEC a2CSS series). Usually
manufacturers provide special program adaptors and PC-software tools. But this programming is also possible by this
measuring receiver with the DiSEqC-Script function.
A set of DiSEqC commands according EN 50494/50607 can be easily edited in a common texteditor end saved with the
ending .dsq. Appropriate files can you get from the manufacturer (e.g. JULTEC). These files must be stored on an USB-
stick which can be connected to the measuring receiver. The script function sends the of DiSEqC commands line by line
over RF-connector to the DiSEqC-Uint.
6.5.8.1 Execute DiSEqC script
First, the device has to be in initial state (running measurements stop by pressing ESC). Plug the USB-Stick containing
the Script-Files to the receiver and connect the the DiSEqC-Uint via RF-Connector.
Selecting
LNB
> DiSEqC Script, choose one of the “.dsq-files” shown in a directory and by pressing
ENTER
the function
starts. A window shows the first comment of the script file and the currently processed line. In the last line of the
window a possible DiSEqC-Responce is shown.
6.5.8.2 Skript structure
Sample of a DiSEqC-Script:
WideBand mode UAS 474 2 x 950..3000 MHz
E2 7F FB 40 00 00 03 52 00 00 03 52; LOF 9,75 GHz, IF 950..3000 MHz
E2 7F FB FE; store data
// comment