2-2
■
When moving the centrifuge, lift up the bottom of the centrifuge
body by two or more persons using equal forces to move the
centrifuge to the desired place.
■
Do not drop the centrifuge, otherwise damage or injury may occur.
CAUTION
Prepare a power supply that meets the following conditions.
1. The power supply voltage must be the same as that indicated on the
nameplate of the centrifuge and the voltage variation must be within
the range given in Table 2-1.
2. The current capacity must be more than given in Table 2-1.
3. In connecting the cord, install a knife switch or circuit breaker of given
value in Table 2-1.
When the centrifuge is connected to an outlet by means of a plug, use
a 3-pin plug with a ground terminal.
4. Use single-phase power.
5. The outlet must have a ground terminal and its ground resistance must
be less than 100 ohm.
Table 2-1 Rated Voltage, Current and Acceptable Voltage Range and Current Requirement
WARNING
●
The power cable alone should be connected to the plug socket.
●
Do not use a branched plug socket, which may cause overheating or fire.
WARNING
Rated Voltage
100-115V
200-240V
Acceptable Voltage Range
90V to 126V
180V to 264V
Current requirement
8A
5A
2-4
.
Power Requirement
Section 2 Installation and Power Supply