2-1
Contd. on next page.
●
Never move the centrifuge while the rotor is rotating or while the rotor is
attached to the centrifuge. Otherwise, the drive shaft may become bent
or the rotor and the bucket may come off, resulting in an accident or
damage to the centrifuge.
●
Ensure that rotor and bucket are removed from the centrifuge and that
the power cord is disconnected from the wall socket.
●
Moving the position of the centrifuge while the power is turned on may
cause electrification accident or functional failure of the centrifuge.
WARNING
When installing this centrifuge, keep 3meters or more away from AM radio.
Noise generated by this centrifuge causes the sound of AM radio deteriorated.
A clearance of 30 cm minimum must be provided around the centrifuge.
If the centrifuge is driven into uncontrolled rotations due to a failure, secondary
damage can result from energy absorbed by the rotation.
WARNING
Do not install the centrifuge on an inclined, slippery, or unstable surface.
Violent vibration may occur.
CAUTION
Section 2
Installation and Power Supply
2-1.
Unpacking
When the centrifuge is taken out of a corrugated carton box, check the following.
(1)
Upon receiving the centrifuge, examine it for any visible damage caused during
transportation.
If any is found, contact the dealer immediately.
(2)
Confirm that all the accessories listed in [7 - 2. Standard Accessories] are included with
the delivery.
Refer to page 7-1.
2-2.
Place of installation
2-3.
Movement of centrifuge