Active output
short-circuit-proof
Terminals 4.1/4.2 :
for electromechanical (EMC) or
electronic (EC) totalizers
Terminals 4/4.1/4.2:
for electronic (EC) totalizers
Amplitude
:
approx. 27 V
Load rating
see Table "pulse width"
Passive output
Terminals 4/4.1
: open collector for connection of
active electronic totalizers (EC)
or switchgear
Input voltage
: 5 to 30 V
Load current
: max. 100 mA
Pulse width is automatically choosen, depending on actual out
frequency.
Frequency f at Q=100%
Pulse width
Load rating of active output
Load current
Load
0.0028 Hz < f
≤
1 Hz
500 ms
≤
150 mA
≥
180 Ohm
approx.
50% duty
1 Hz < f
≤
10 Hz
cycle (1:1)
≤
25 mA
≥
1 k Ohm
50% duty
10 Hz < f
≤
1000 Hz
cycle (1:1)
≤
25 mA
≥
1 k Ohm
1000 Hz < f
≤
2547 Hz
160 µs
≤
25 mA
≥
1 k Ohm
2547 Hz < f
≤
10000 Hz 50 µs
≤
25 mA
≥
1 k Ohm
45