ZERO SET
:
Zero point calibration.
Select between the factory set default value for the zero point
or perform a zero point calibration for the actual situation.
See paragraph C.1.4 for the precise conditions and instruc-
tions for a calibration.
C.2.3
Plausibility check
After any changes are made to the parameter settings the
microprocessor will perform a plausibility check on the
changed settings, before the measuring mode can be acti-
vated. If no "impossibilities" in the configuration are detected
the microprocessor asks for confirmation before the new set-
tings are stored.
If discrepancies in the parameters entered are found the
microprocessor will display the message "PARAMERROR"
which prevents you from storing parameters that are incor-
rect; the options available to correct the configuration are
described in section D.6.
The next function which will appear is again, the first function
of this Start Up Menu: "DIAMETER".
The Start Up Menu is left by pressing the key once from
any function name in the Start Up Menu. When doing this,
the Main Menu will be entered (function 1.0.0.)
34
ALTOSONIC
Operation
UFM 600 T