C.7
Trouble Shooting
1
The flowmeter does not respond after connection to
the main power supply
(no display or output messages).
- Check fuse.
2
The error marker * appears on the display
-
Leave the measuring mode and enter the Quit/Reset
menu
(press ; see section C.5).
Cure the cause if necessary, and select
ERR.ACQUIT to reset the error marker.
3
Low signal marker: 3 or 4 markers on.
Cause
Solution
Incorrect transducer
Reposition transducers (see sec-
tion position C.1.3) using sufficient
coupling grease and check the
transducer distance afterwards.
Poor sonic coupling
Check the pipe wall surface;
between pipewall and
remove any rust, dirt and/or thick
transducer surface
layers of paint or coating.
Apply sufficient grease on the
transducers and make sure not to
wipe off any grease from the trans-
ducers while they are being rein-
stalled.
Pipe is not completely
Fill the pipe completely or select a
filled
pipe section that is always com-
pletely filled with liquid.
Broken signal cable
Use a multimeter to check the
condition of the cables and the
connectors. See your supplier for
46
ALTOSONIC
Operation
UFM 600 T