7.30830.32.00
41
Factors for volume F
M
Volumetric unit
TEXT VOL.
Factor F
M
Setting
Cubic meters
m
3
1.0
1.00000 E+0
Liter
Liter
1000
1.00000 E+3
Hectoliters
h Lit
10
1.00000 E+1
Deciliters
d Lit
10000
1.00000 E+4
Centiliters
c Lit
100000
1.00000 E+5
Milliliters
m Lit
1000000
1.00000 E+6
US gallons
USGal
264.172
2.64172 E+2
Millions US gallons
USMG
0.000264172
2.64172 E-4
Imperial gallons
GBGal
219.969
2.19969 E+2
Mega imperial gallons
GBMG
0.000219969
2.19969 E-4
Cubic feet
Feet3
35.3146
3.53146 E+1
Cubic inches
inch3
61024
6.10240 E+4
US barrels liquid
US BaL
8.36364
8.38364 E+0
US barrels ounces
US BaO
33813.5
3.38135 E+4
Factors for time F
T
Time unit
TEXT TIME
Factor F
T
Setting
Seconds
sec
1
1.00000 E+0
Minutes
min
60
6.00000 E+1
Hours
hr
3600
3.60000 E+3
Day
DAY
86400
8.64000 E+4
Year
YR
31536000
3.15360 E+7
5.15
F/R mode, forward/reverse flow measurement
•
Refer to Section 2.5.6 for electrical connection of outputs.
•
Define direction of forward flow,
see Fct. 3.02, subfunction “FLOW DIR.”:
in conjunction with F/R operation, this is where to set the direction of the forward flow.
“+” means the same direction as shown by the arrow on the primary head.
“-“ means the opposite direction.
•
Set one of the
status outputs
to “SIGN I”, “SIGN P” or “SIGN P2”, see Fct. 1.08-1.10 (1.07).
Dynamic behavior of outputs in case of “SIGN I, P or P2” see Section 5.9.
•
Current and/or pulse outputs
must be set to “2 DIR.”, see Fct. 1.05, 1.06 and 1.07,
subfunctions “FUNCT. I”, “FUNCT. P” and “FUNCT. P2”.