Characteristic of status outputs
Switch open
Switch closed
OFF
(switched off)
no function
ON
(e.g. operation indicator)
power supply OFF
power supply ON
SIGN I
(F/R mode)
forward flow
reverse flow
SIGN P/P2
(F/R mode)
forward flow
reverse flow
FULL SCALE
(full-scale indicator)
inactive
active
AUTO. RNG
. (automatic range
change)
high range
low range
OVERFL. I
(I range exceeded)
current output O.K.
current output range exceeded
OVERFL. P/P2
(P range exceeded)
pulse output O.K.
pulse output range exceeded
ALL ERRORS
error
no error
FATAL. ERRORS
error
no error
INVERS A1
: Status output A2…
when A1 is closed
when A1 is open
INVERS D1
: Status output D2…
when D1 is closed
when D1 is open
EMPTY PIPE
(empty tube
identification option)
when measuring tube is empty
when measuring tube is full
For factory settings please refer to sect. 2.5.7.
5.10
Control inputs C1 and C2
Fct. 1.11 CONTROL C1
and/or
Fct. 1.12 CONTROL C2
Press
→
key
Press
→
key
Select function for the control inputs,
press
↑
or
↓
key
•
OFF
switched off, no function
•
OUTP. HOLD
hold output values
Function also acts on display and totalizer
•
OUTP. ZERO
set outputs to “min. values”
Function also acts on display and totalizer
•
TOTAL. RESET
reset totalizer
•
ERROR. RESET
acknowledge/delete error messages
•
EXT. RNG.
external range change for automatic range change, see sect. 5.20.
Settling range 5 – 80 PERCENT = low to high ratio 1:20 to 1:25, value must be greater than
that of Fct. 1.03 L.F. CUTOFF.
Press
→
key to change to numerical setting, 1st number (cursor) flashes.
Change flashing number (cursor) with keys
↑
and
↓
. Use
←
and
→
keys to shift cursor 1 place to left or right.
Flashing numbers can also be directly set with the 10-key keyboard.
Press
↵
key to return to Fct. 1.11 CONTROL C1 or to Fct. 1.12 CONTROL C2.
Please refer to sect. 2.5.7 for factory settings.
Connection diagram see sect. 2.5.6.
5.11
Language
Fct. 3.01 LANGUAGE
Press
→
key
Select language for texts in display
•
D
German
•
GB/USA
English
•
F
French