5.9
Status outputs A1 / A2 and D1 / D2
PLEASE NOTE:
Connection diagrams see sect. 2.5.6.
Status outputs
A1
A2
D1
D2
Select
Fct.
_ . _ _
then press
→
key
1.07
1.08
1.09
1.10
Terminals
A1 / A
⊥
A2 / A
⊥
D1 / D
⊥
D2 / D
⊥
Max. switching
current
•
100 mA (AC or DC)
•
200 mA (DC polarized)
see sect. 6.3
100 mA (AC or DC)
100 mA (AC or DC)
100 mA (AC or DC)
Remark
“STATUSOUTP.” must
be selected in Fct. 3.07
HARDWARE,
subfunction
“TERMINALS”.
-
-
-
PLEASE NOTE:
Select function for status outputs,
press
→
key
•
ALL ERROR
indicate all errors
•
FATAL.ERROR
only indicate fatal errors
•
OFF
switched off, no function
•
ON
signals the operation of the flowmeter
•
SIGN I
F/R mode
•
SIGN P/P2
F/R mode
•
OVERFL. I
Exceeding output ranges
•
OVERFL. P/P2
Exceeding output ranges
•
INVERS. A1
switches output A2 inverse to A1. A1 and A2 then operate as change-over elements with common
centre grounding contact A
⊥
. Only available when status output is selected in Fct. 3.07 “TERM.
A1”.
•
INVERS. D1
switches output D2 inverse to D1. D1 and D2 then operate as change-over elements with common
centre grounding contact D
⊥
.
•
EMPTY PIPE
signals that measuring tube is empty, only with option “empty tube detection”
•
AUTO. RNG.
Automatic range change. Setting range 5 – 80 PERCENT (=high to low ratio, 1:20 to 1:25, value
must be higher than that of Fct. 1.03 L.F. CUTOFF).
•
FULL SCALE
Select flow direction
(characteristic) for full-scale range
•
+ DIR.
•
- DIR.
•
2 DIR.
Select with
↑
and
↓
keys
.
Define full-scale range
XXX - YYY
0-150%
0-150%
normally open contact:
XXX > YYY
normally closed contact:
XXX < YYY
hysteresis:
difference between XXX and YYY
Press
→
key to change to numerical setting, 1st number (cursor) flashes.
Change flashing number (cursor) with keys
↑
and
↓
. Use
←
and
→
keys to shift cursor 1 place to left or right.
Flashing numbers can also be directly set with the 10-key keyboard.
Press
↵
key to return to Fct. 1.07, 1.08, 1.09 and 1.10 for status outputs A1, A2, D1 and D2.