7.30830.32.00
35
→
→
I ERROR = set error value,
press
→
key
•
22 mA
fixed value
•
0.0 – I
0%
mA
variable value, only variable when I
0%
≥
1 mA, see “RANGE I” above
Select with
↑
and
↓
keys.
Change flashing number (cursor) with keys
↑
and
↓
. Use
←
and
→
keys to shift cursor 1 place to the left or right.
Flashing numbers (cursor) can also be directly set with the 10-key keyboard.
Press
↵
key to return to Fct. 1.05 CUR. OUTP. I
Please refer to sect. 2.5.7 “Standard factory settings”
Refer to sect. 2.5.6 for connection diagrams and to sect. 5.16 for characteristics
.
5.8
Pulse outputs P and A1
Pulse output P
2nd pulse output A1
To be used for …
electronic totalizer
electromechanical or electronic totalizer
Terminals
P and P
⊥
A1 and A
⊥
F
max
at full-scale range Q
100%
10.000 pulses/s
50 pulses/s
F
min
at full-scale range Q
100%
10 pulses/hr
10 pulses/hr
Max. switching current
30 mA (AC or DC)
100 mA (AC or DC)
200 mA (DC polarized)
see sect. 6.3
Remark
-
“PULSOUTP.” must be selected in
Fct. 3.07 HARDWARE, subfunction
“Terminal A1”
Fct. 1.06 PULS P
and/or
Fct. 1.07 PULS2 A1
Press
→
key
Press
→
key
→
→
FUNCT. P = select function for pulse output,
press
→
key
•
OFF
switched off, no function
•
+ DIR.
measurement in one direction, refer to selection of main flow direction in
Fct. 3.02 FLOW METER, subfunction “FLOW DIR.”
•
- DIR.
see + DIR.
•
2 DIR.
2 flow directions, F/R mode, forward/reverse
Select with
↑
and
↓
keys.
Press
↵
key to change to subfunction “SELECT P”.
Exception:
when “OFF” is selected, return to Fct. 1.06 PULS P or Fct. 1.07 PULS2 A1.
→
→
SELECT P = select pulse type,
press
→
key
•
PULSE/VOL.
pulses per unit volume, flow
•
PULSE/TIME
pulses per unit time for 100 % flow
Select with
↑
and
↓
keys.
Press
↵
key to change to subfunction “PULSWIDTH”.