OPERATION
5
37
OPTISENS TSS 2000
www.krohne.com
07/2018 - 4006304401 - OPTISENS TSS 2000 en R01
Step 5: Start calibration
The sensor can be installed in the intended measuring point.
Step 6: Deactivate the hold function
In the upper line of the display product cal.
product cal.
product cal.
product cal. appear. Press
or
until the submenu apply calibration
apply calibration
apply calibration
apply calibration is
highlighted.
•
Press >>>> to enter the chosen menu.
•
Press
or
to answer the question apply calibration?
apply calibration?
apply calibration?
apply calibration? with the following options: Yes
Yes
Yes
Yes or No
No
No
No
•
Press
or
until Yes
Yes
Yes
Yes is shown and press
^
^
^
^
to confirm.
•
please wait
please wait
please wait
please wait appear and a 15 seconds countdown starts.
calibration successful
calibration successful
calibration successful
calibration successful appear
•
Press
^
^
^
^
to confirm.
•
Press
^
^
^
^
several time until the question Save configuration?
Save configuration?
Save configuration?
Save configuration? is highlighted.
•
Press
or
to answer the question Save configuration?
Save configuration?
Save configuration?
Save configuration? with the following options: Yes
Yes
Yes
Yes or No
No
No
No
•
Press
^
^
^
^
to confirm.
INFORMATION!
•
If the calibration was not successful you can choose between load old cal. data
load old cal. data
load old cal. data
load old cal. data and check cal.
check cal.
check cal.
check cal.
data
data
data
data. If you choose load old cal. data
load old cal. data
load old cal. data
load old cal. data the old calibration data will be loaded and the current
data will be deleted. If you choose check cal. data
check cal. data
check cal. data
check cal. data the actual data will be saved and you can
check these data.
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level.
•
Press
or
until the main menu quick setup
quick setup
quick setup
quick setup is highlighted.
MAIN MENU
> A quick setup
> A quick setup
> A quick setup
> A quick setup
B test
C setup
D service
•
Press >>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display quick setup
quick setup
quick setup
quick setup appears, beneath the
submenu language
language
language
language is highlighted.
•
Press
or
until the submenu manual hold
manual hold
manual hold
manual hold is highlighted.
•
Press >>>> to enter the chosen menu.
You are on the second submenu level. In the upper line of the display quick setup
quick setup
quick setup
quick setup appears, beneath the
option on
on
on
on is highlighted.
•
Press
or
to choose the option off
off
off
off.
•
Press
^
^
^
^
to confirm.
•
Press
^
^
^
^
to get back in the submenu.