3.3 Storage and transport
•
Store the device in a dry, dust-free location.
•
Avoid continuous direct sunlight.
•
The original packing is designed to protect the equipment. It has to be used if the device is
transported or sent back to the manufacturer.
3.4 Configuration of a measuring point
The sensor has to be operated with an external signal converter.
A complete measuring point consists of at least two or three parts:
•
Signal converter
•
OPTISENS TSS 2000 sensor (fixed cable)
•
Immersion assembly or other adequate housing
3.5 Installation procedure
The sensor is factory calibrated and ready to use. However, it is recommended to calibrate the
sensor with one or more process samples before it is installed into its final measuring location.
To install the device in the correct way, follow the order of the following sections and their
instructions.
1. Connect the sensor to the signal converter.
2. Calibrate the sensor.
3. Mount the sensor into the immersion assembly. (For further information refer to the manual of
the assembly)
4. Install the sensor into its final measuring location.
CAUTION!
•
Never touch or scratch the measurement windows of the sensor.
•
Make sure that the measurement windows are clean and dust-free. If necessary, clean the
measurement windows as described on page 40
.
•
The device must not be heated by radiated heat (e.g. exposure to the sun) to a electronics
housing surface temperature above the maximum permissible ambient temperature. If it is
necessary to prevent damage from heat sources, a heat protection (e.g. sun shade) has to be
installed.