5
OPERATION
36
OPTISENS TSS 2000
www.krohne.com
07/2018 - 4006304401 - OPTISENS TSS 2000 en R01
Step 4b: Calibration (non linear)
INFORMATION!
Make sure there are no air bubble in front of the optics as they will be causing measurement
signals due to reflection and light scattering effects and such might cause a wrong reading.
In the upper line of the display product cal.
product cal.
product cal.
product cal. appear. Press
or
until the submenu calibration mode
calibration mode
calibration mode
calibration mode is
highlighted.
•
Press >>>> to enter the chosen menu.
Press
or
until non linear
non linear
non linear
non linear is highlighted.
•
Press
^
^
^
^
to confirm.
Press
or
until take sample 0
take sample 0
take sample 0
take sample 0 is highlighted.
•
Press >>>> to enter the chosen menu.
•
Press
or
to answer the question take sample?
take sample?
take sample?
take sample? with the following options: Yes
Yes
Yes
Yes or No
No
No
No
•
Press
^
^
^
^
to confirm.
The actual ADU value is shown
•
Press
^
^
^
^
to confirm.
•
Press
or
to answer the question store cal. value?
store cal. value?
store cal. value?
store cal. value? with the following options: Yes
Yes
Yes
Yes or No
No
No
No
•
Press
^
^
^
^
to confirm.
Information! The calibration process for take sample 1,2
take sample 1,2
take sample 1,2
take sample 1,2 is equal to take sample 0
take sample 0
take sample 0
take sample 0.
Press
or
until TSS value 0
TSS value 0
TSS value 0
TSS value 0 is highlighted.
•
Press >>>> to enter the chosen menu.
In the upper line of the display TSS value
TSS value
TSS value
TSS value appear. Press
or
until the submenu TSS value 0
TSS value 0
TSS value 0
TSS value 0 is
highlighted.
•
Press >>>> to enter the chosen menu.
Press
or
to change the value and press >
>>> to move to the next digit in the displayed value.
Insert the determine laboratory value (e.g. 0.000 g/l) and confirm with
^
^
^
^
.
•
Press
or
to answer the question store value?
store value?
store value?
store value? with the following options: Yes
Yes
Yes
Yes or No
No
No
No
•
Press
^
^
^
^
to confirm.
•
Press
^
^
^
^
to get back in the submenu.
Information! The calibration process for TSS value 1,2,
TSS value 1,2,
TSS value 1,2,
TSS value 1,2, is equal to TSS value 0
TSS value 0
TSS value 0
TSS value 0.
INFORMATION!
In case that TSS values are entered later, do not start the calibration by apply calibration
apply calibration
apply calibration
apply calibration before
all TSS values were stored. Work with the previous calibration until all lab TSS values are
available and can be entered.