C5.3 1.meas. page
This menu groups all settings for the first measuring page. The menus are
equal for the first and second measuring page.
C5.3.1 function
This function allows to specify the
number of lines which display the
measured value(s). Options:
•
one line: display shows one line of
measuring results on this page.
•
two lines: display shows two lines
of measuring results on this page.
•
three lines: display shows two
three of measuring results on this
page.
C5.3.2 measurement 1.line
This function allows to define the
value which activates the display.
The available options are TSS value
and temperature.
C5.3.3 range
Defines the min. and max. limits
TSS range: 0.000...50.00 g/l
Temp. range: -10.0...+200
°
C /
14...392
°
F
C5.3.4 limitation
Defines the min. and max. limits
Range: -150.0...+150%
C5.3.5 time constant
Defines the min. and max. limits
Range: 0.001...100 seconds
C5.3.6 format 1.line
This function allows to define the
number of digits after the decimal
point (i.e. the decimal places).
Options:
•
automatic
•
X. up to X.XXXX
If one line is selected in C5.3.1 the
display shows maximum X.XXX.
If two or three lines are selected in
C5.3.1 the display shows maximum
X.XXXX.
C5.3.7 measurement 2.line
This function allows to define which
measured value appears in the
second line of the display. The
existence and structure of this menu
within your device depends on the
hardware setting and the kind of
sensor connected to the process
input.
Options:
•
bargraph
•
operating hours
•
TSS value
•
temperature
C5.3.8 format 2.line
These function allows to define the
number of digits after the decimal
point (i.e. the decimal places). The
existence of this menu depends on
the hardware setting, the options
comply with the options for the
format of the first line of the display.
Options:
•
automatic
•
X. up to X.XXXXXXXX
Function
Settings / descriptions