INSTALLATION
3
15
OPTISENS TSS 2000
www.krohne.com
07/2018 - 4006304401 - OPTISENS TSS 2000 en R01
3.7 Mounting the sensor into an assembly
For further instructions on installation into an immersion assembly refer to the assembly
manual.
Use an assembly that does not fix the sensor cable or require the sensor to be screwed into the
assembly. If the sensor needs to be screwed for mounting or dismounting make sure that the
sensor cable is turned into the same direction.
Calibrate the sensor before installing it into the assembly. For further information refer to
Calibration
on page 32.
Installing procedure
•
Insert the sensor cable through the immersion assembly.
•
Fasten the sensor to the tip of the telescopic rod by tighteninng the screws of the 2 holding
plates.
•
Connect the wires to the MAC 100 signal converter. For further information refer to
Connecting the sensor cable to the signal converter
on page 16.
For removing the sensor, repeat the steps above in reverse order.
DANGER!
All work on the electrical connections may only be carried out with the power disconnected.
CAUTION!
Do not turn the cable gland on the sensor this might cause a sensor leak and damage the
electronics inside. While mounting or dismounting the sensor, the senor cable must not be fixed
or trapped as this might loosen the water tight gland connection from the sensor.