Maintenance
93
9.4 Maintenance work
Maintenance work
Description
Cleaning the humidifier
1
Unscrew the humidifier from the angle connection
2
Unscrew the lid from the humidifier and pour out any
water residues
3
Clean the humidifier with clear, warm water and disinfect
the humidifier as instructed
Or alternative
4
Disinfect humidifier and lid thermally in the dishwasher.
Use, if existing, the rinse cycle for glasses.
5
Fill fresh, distilled water into the humidifier
6
Turn the lid back on
7
Fasten the humidifier on the angle connection
NOTE!
When changing patients the humidifier must be
disposed of.
Sterile water system
If the
aeroplus 5
is used with a sterile water system, the
following must be observed:
– Sterile water systems must not be cleaned and refilled.
– The old water flask must be disposed of.
– The information provided by the sterile water producer
(package insert) must be observed!
Cleaning the nasal
cannula
1
Disconnect the hose of the nasal cannula from the
aeroplus 5
.
2
Clean the nasal cannula in warm soapsuds. You may
alternatively use a weak acetic solution (10% vinegar,
90% water).
3
Rinse the nasal cannula with lots of clear water.
4
Let the nasal cannula dry in air. The nasal cannula may
only be used again for the therapy after it has properly
dried.