aeroplus 5
Safety
67
ATTENTION! Keep the minimum distance!
The air intake of the
aeroplus 5
is located on the back side of the unit, the
following is therefore of importance:
–
Set up the
aeroplus 5
oxygen concentrator in a well aerated place.
–
There should be a minimum distance of 30cm from walls, curtains and
other large objects (e.g. cupboards), so that an unrestricted entry of air is
assured at the back of the unit.
–
The
aeroplus 5
oxygen concentrator must not be used directly beside or
even stacked with other equipment.
ATTENTION! Avoid overheating!
The unit is air cooled, in order to avoid overheating it must not be placed near
heaters, etc.
ATTENTION! Air pollution!
Set up the
aeroplus 5
oxygen concentrator only in places without air pollution or
smoke.
3.5 What to do if a hose caught fire
Should a hose catch fire despite all protective measures, it is not enough to just switch the
unit off, because oxygen will continue to flow for a while after the unit has been switched off.
Take the following steps:
–
Pull the oxygen hose off the unit to interrupt the oxygen supply.
–
Choke the flames (e.g. with a blanket).
–
Ventilate well after extinguishing the fire, because a burning PVC-hose will emit
toxic gases.
The metal connection on the oxygen outlet acts as a fire inhibitor, preventing the flames from
spreading into the unit.