aeroplus 5
Taking into service
78
Connection to the nasal cannula
5
Plug the FireSafe Nozzle in oxygen
flow direction into the nasal cannula.
ATTENTION!
Observe flow direction indicated
by the arrow!
ATTENTION!
For the best protection of the
patient, place the FireSafe Nozzle
in the oxygen tubing routing as
close to the patient as possible!
Connection with the O2 tubing
6
Connect the other terminal to the
short oxygen tubing (0.33 m).
7
Connect the oxygen tubing to the
humidifier.
NOTE!
The humidifier can be used directly
at the oxygen concentrator or with
the kit for external humidification
(near to the patient).
WARNING! Danger of tripping over!
On the back of the
aeroplus 5
you will find a winding device for the mains lead.
This should be used when the unit is out of use, to avoid any danger of tripping
over.
Nasal cannula
Nasal cannula
O2 tubing