![Krober aeroplus 5 Instructions For Use Manual Download Page 86](http://html1.mh-extra.com/html/krober-0/aeroplus-5/aeroplus-5_instructions-for-use-manual_2004043086.webp)
aeroplus 5
Operation
86
Alarm category
Description
Mains failure
LEDs: yellow flashing
Tone: Medium priority
Possible Cause:
The power supply for the unit is interrupted. This causes
immediate function failure of the Kröber
aeroplus 5
!
Countermeasures:
The following should be checked:
– Is the mains lead properly plugged into the socket?
– Has a fuse tripped? Check the fuse, replace if necessary.
Note
If a function test concerning the mains failure alarm is to be
performed, you should proceed as follows:
– Pull the mains plug out of the socket.
– Switch on the unit.
Mains failure
LEDs: off
Tone: High priority
Possible Cause:
The power supply for the unit is interrupted. This causes
immediate function failure of the Kröber
aeroplus 5
!
Countermeasures:
The following should be checked:
– Is the mains lead properly plugged into the socket?
– Has a fuse tripped? Check the fuse, replace if necessary.
Note
If a function test concerning the mains failure alarm is to be
performed, you should proceed as follows:
– Pull the mains plug out of the socket.
– Switch on the unit.