![Krober aeroplus 5 Instructions For Use Manual Download Page 80](http://html1.mh-extra.com/html/krober-0/aeroplus-5/aeroplus-5_instructions-for-use-manual_2004043080.webp)
aeroplus 5
Taking into service
80
1
4 Fill the humidifier
up to the top MAX.-
mark
(1)
.
NOTE!
Only use the supplied or identical
humidifier with O/N 48-7005.
1
3
2
5 Screw the angle adapter (1)
to the
humidifier.
6 Plug the safety hose for oxygen
(extension hose)
onto
the
angle
adapter (2)
.
7 Push the short 0.33 m oxygen safety
tubing
with its socket end over the
outlet connection
of the
oxy 5000
(3)
.
Connection to the nasal cannula
8
Plug the FireSafe Nozzle in oxygen
flow direction into the nasal cannula.
ATTENTION!
Observe flow direction indicated
by the arrow!
ATTENTION!
For the best protection of the
patient, place the FireSafe Nozzle
in the oxygen tubing routing as
close to the patient as possible!
Nasal cannula