8.3.2 Alarm categories
Alarm category
Description
Temperature
LED: yellow flashing
Tone:
Medium priority
Possible Cause:
The operating temperature inside the concentrator is too high.
Countermeasures:
– The unit should be switched off immediately.
– Check whether the air flow into the unit is restricted. Also
make sure that the unit has a sufficient clearance to other
objects (wall, cupboard, etc.).
– It may be necessary to replace the coarse dust filter in the
back wall of the unit. Further information concerning this
matter can be found in the section "Maintenance".
NOTE!
The oxygen supply is immediately stopped to protect the
patient. However, the compressor keeps on running.