DIAGNOSTICS
Using the Diagnostics
10DEC05
DG4825-1
Page
23 of 180
[System Self Test]
Important
For the test procedure, see
Completes the following diagnostic tests:
•
“GALVO BOARD COMMUNICATION” - Checks the
connection of the CABLE between:
–
GALVO BOARD A4
–
MASTER CENTRAL PROCESSING UNIT BOARD
A2
•
“GALVO REFERENCE” - Reads the GALVO AGC
CHANNEL connected to the reference voltage of the
GALVO.
•
“GALVO MOTOR” - Reads the voltage of the GALVO
AGC to check for problems with the GALVO.
•
“GALVO SWEEP” - Checks the functions of reading and
writing of the REGISTERS on the GALVO BOARD A4.
•
“LASER OFFSET”
•
“PMT BOARD TEST” - Checks the connection of the
CABLE between:
–
MASTER CENTRAL PROCESSING UNIT BOARD
A2
–
PMT/DAS BOARD A5
•
“MASTER CENTRAL PROCESSING UNIT BOARD
MEMORY TEST” - Checks for errors in the reading and
writing of data to each of the image buffers.