DIAGNOSTICS
Error Codes
10DEC05
DG4825-1
Page
64 of 180
10022 - MSC BOARD A1: Software malfunction
10024 - MSC BOARD A1: Software malfunction
10029 - READER detected the CASSETTE BAR CODE label as backward
10030 - READER did not read the CASSETTE BAR CODE
Description: The DRIVE ROLLER received a signal when waiting for an interrupt or
timeout to occur.
Cause 1: The software for the MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
Install the software again for the MOTION SYSTEM CONTROL BOARD A1.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
Do the checkout for the
MOTION SYSTEM CONTROL BOARD A1
Description: A timeout occurred setting up the motion for the DRIVE ROLLER.
Cause 1: The software for the MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
Install the software again for the MOTION SYSTEM CONTROL BOARD A1.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
Do the checkout for the
MOTION SYSTEM CONTROL BOARD A1
Cause 1: The CASSETTE was inserted backward into the CR 825/850 SYSTEM.
Action:
Insert the CASSETTE correctly.
Cause 2: The BAR CODE LABEL is installed backward on the CASSETTE.
Action:
1.
Remove the BAR CODE LABEL.
2.
Install a new BAR CODE LABEL.
Cause 1: The CASSETTE:
•
has a bar code that you cannot read
•
does not have a bar code
Action:
Install the correct bar code.