DIAGNOSTICS
Error Codes
10DEC05
DG4825-1
Page
113 of 180
47520 - 47522 - Internal communications failure to UPS1
47550 - The BATTERY in UPS1 is too low. System will be shutdown in 1
minute
47551 - Unable to establish communications with UPS1
47950 - Display only - no Log Message
47951 - Display only - no Log Message
Cause:
Communication error.
Action 1:
1.
Check the RS-232 CABLE to COM3.
2.
Do the checkout for the
UNINTERRUPTED POWER SUPPLY UPS1
.
Cause 1: Malfunction of the utility power.
Action:
Check the POWER PLUG and wait for the power to begin.
Cause 2: The level of the BATTERY is 25% or less of the maximum level.
Action:
Do the checkout for the
UNINTERRUPTED POWER SUPPLY UPS1
.
Cause 1: Malfunction of the utility power.
Action:
Check the POWER PLUG and wait for the power.
Cause 2: The level of the BATTERY is 25% or less of the maximum level.
Action:
Do the checkout for the
UNINTERRUPTED POWER SUPPLY UPS1
.
Description: The UNINTERRUPTED POWER SUPPLY UPS1 detects a utility power
failure condition that did not clear within 20 seconds.
Cause:
Malfunction of the utility power.
Action:
Check the POWER PLUG.
Description: The UNINTERRUPTED POWER SUPPLY UPS1 detects that the
BATTERY charge low is indicated from the UNINTERRUPTED POWER
SUPPLY UPS1.