DIAGNOSTICS
Checkout Procedures
10DEC05
DG4825-1
Page
146 of 180
2
If the measurements are not correct, do the checkout for the
Important
, you must:
•
install the INTERLOCK KEY, or
•
close the FRONT DOOR
3
Check the voltages at the test points:
4
If the voltage is not correct, do the checkout for the
Important
To check the +10 V DC, do the adjustment for the GALVO BOARD A4.
5
Check the voltages at the test points:
6
If any of the voltages do not match the table, install a new GALVO BOARD A4.
MULTIMETER
Voltage
Ground
Positive
TP1
TP16
–24
±
0.5 V DC
MULTIMETER
Voltage
Negative
Positive
TP1
TP5
+15
±
0.25 V DC
TP1
TP6
−
15
±
0.25 V DC
TP1
TP13
+10
±
0.1 V DC
TP1
TP15
+5
±
0.1 V DC