DIAGNOSTICS
Using the Diagnostics
10DEC05
DG4825-1
Page
13 of 180
Table 2
Calibration
BUTTON
Description
[View Collector Profiles]
Important
For the procedure, see
Calibration for the CR 825/850 SYSTEM
Displays the COLLECTOR PROFILE for each size of CASSETTE,
including:
•
number of steps from “HOME” to “START OF SCAN”
•
“OFFSET” - Number of pixels from “START” to the diagnostic
image
•
“Amplitude” - Number of pixels from “OFFSET” to the last pixel
of the diagnostic image
•
PMT 1 - 2: Correction value of each PMT for gain adjustment
[Scan Calibration]
Caution
•
All COVERS must be installed when using this option.
•
For the procedure, see
Calibration for the CR 825/850
•
Completes the calibration of:
–
GALVO
–
PMT 1
–
PMT 2
•
Starts a COLLECTOR PROFILE for all sizes of STORAGE
PHOSPHOR PLATES.
•
Allows calibration of the PLATES exposed at 20 mR
± 5
mR.
•
Displays the date and time of the last successful calibration.