43
Configuration
Menu item
Action
Choices
TC-LIQUID / CIP / SIP
Temp compensation,
process medium
For pH measurement
only: Enter temperature
compensation of the
process medium.
Enter value using
keys.
Press enter to confirm.
Cleaning cycles
on/off
MEMO SENS
Select ON or OFF using
keys.
Activates/deactivates log-
ging in extended logbook
Press enter to confirm.
ON/OFF
Sterilization cycles
on/off
MEMO SENS
Select ON or OFF using
keys.
Activates/deactivates log-
ging in extended logbook
Press enter to confirm.
ON/OFF
The cleaning and sterilization cycles are counted to measure the load
on the sensor.
Suitable for biochemical applications (process temp approx. 0...50 °C,
CIP temperature > 55 °C, SIP temperature > 115 °C).