45
Configuration
Menu item
Action
Choices
Select measuring
mode
MEMO SENS
Select measuring mode
using
keys.
DO:
Measurement in liquids
GAS:
Measurement in gases
Press enter to confirm.
dO %,
dO mg/l
dO ppm
GAS %
Calibration medium:
air/water
MEMO SENS
Select calibration
medium using
keys.
AIR: Cal medium air
WTR: Cal medium water
Press enter to confirm.
CAL_AIR
CAL_WTR
Calibration timer
MEMO SENS
Select/deselect
calibration timer using
keys.
Press enter to confirm.
OFF
ON
(ON: Calibration
cycle)
MEMO SENS
Enter calibration cycle in
hours using
keys.
Press enter to confirm.
0...9999 h
0168 h
Note for the calibration timer:
When Sensocheck has been activated in the Configuration – Alarm menu,
the expiration of the calibration interval is indicated by Sensoface:
Display
Status
Over 80 % of the calibration interval has already past.
The time remaining until the next due calibration can
be seen in the diagnostics menu (see "Diagnostics").
The calibration interval has been exceeded.