95
Zero calibration
Display
Action
Remark
MEMO SENS
Select calibration,
press enter to proceed.
Select Sensor_OXY,
press enter to proceed.
Select CAL ZERO,
press enter to proceed.
MEMO SENS
Ready for calibration.
Hourglass blinks.
Place sensor in oxygen-
free medium
Display (3 sec)
Now the device is in
HOLD mode.
MEMO SENS
Secondary display:
Sensor current
measured
Main display:
Zero current.
Press enter to save this
value or correct using
arrow keys and then
save by pressing enter.
MEMO SENS
Display of slope
Display of new zero current.
Press enter to end calibra-
tion, place sensor in process
Sensoface display
MEMO SENS
The oxygen value is
shown in the main
display, “enter“ blinks.
Stop Hold by pressing
enter.
New calibration:
Select REPEAT,
press enter key.
MEMO SENS
Quit by pressing enter. After end of calibra-
tion, the outputs re-
main in HOLD mode
for a short time.